Class Description


Class for controlling the Power Supply of Delta Elektronka SM 70-22.

Families: PowerSupply

Key words:

Platform: Unix Like

Language: Cpp

License:

Contact:

Hardware


Manufacturer: Delta Elektronika

Product:

SM70-22

Bus: Ethernet

Class interface


Attributes:

Name Description
EmissionCurrentScalar: DevDouble Electron emission current. Set to 0 for turning the filament off.
CMEBiasVoltageScalar: DevLong Electron multiplier bias high voltage. If set to 0, it turns off the Electron Multiplier and enable Faraday Cup detection.
NoiseFloorScalar: DevLong Electrometer's Noise-floor. Set the rate and detection limit for ion current measurements.
FinalMassScalar: DevLong Final Mass value for Analog and Histogram scans.
InitialMassScalar: DevLong Initial mass value for Analog and Histogram scans.
SingleMeasurementOutputScalar: DevLong Ion current from a measurement
StepsPerAmuScalar: DevLong Number of steps executed per amu for analog scan.
TotalIonCurrentScalar: DevDouble Total ion current measured at the end of each scan.
IonCurrentsSpectrum: DevLong Ion currents from scan.

Commands:

Name Description
StateInput: DevVoid
Output: State
Device state
This command gets the device state (stored in its device_state data member) and returns it to the caller.
StatusInput: DevVoid
Output: ConstDevString
Device status
This command gets the device status (stored in its device_status data member) and returns it to the caller.
WriteReadSocketInput: DevString
Output: DevString
Write to the socket and get answer.
InitHardwareInput: DevVoid
Output: DevVoid
Initialize communications and check the ECU hardware.
SwitchOffInput: DevVoid
Output: DevVoid
Initialize communications and ECU hardware, reset the RGA to its factory default settings, turns off the filament and the CDEM.
SingleMassMeasurementInput: DevLong
Mass setting (mass to charge ratio in amu units) Possible values: 1-200. If 0, the RF/DC voltages are completely shut down and not measurement is performed (no ion current is transmitted back), use it at the end of a set of single mass measurements to make sure the RF/DC are completely turn off.
Output: DevVoid
Performs a single mass measurement at a mass setting given as argument. The otuput will be written in the attribute: SingleMeasuremetOutput.
StartMultipleScansInput: DevLong
Number of scans.
Output: DevVoid
Execute one of multiple analog scans under the present scan conditions.
StopScanningInput: DevVoid
Output: DevVoid
Stop continuos or serie of scans.

Pipes:

Properties:

Name Description
PortNbDevLong Port Number for connecting to the power supply
HostnameDevString Hostname of the power supply.
SimulationModeDevULong 0 -> real mode, 1 -> simulation mode

23 Feb 2017, Piotr Goryl
Updated:
The device class has been updated.
You can see previous version here .



19 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You are looking at this version now.



2 Jan 2017, Piotr Goryl
Created:
The device server has been added to catalogue.