This is historical information of device classes implemented in FPS3010LCtrl device server.

Use this link to find the valid information.

Development status: Released
Information status: New

Contact:



Class Description


Tango device server for Attocube FPS3010 interferometer

Families: MeasureInstruments

Key words:

Platform: Unix Like

Language: Cpp

License: GPL

Contact:

Hardware


Manufacturer: Attocube

Product:

Bus: Not Applicable

Class interface


Attributes:

Name Description
DeviceFeaturesScalar: DevLong Bitfield: 0x01: Ethernet enabled 0x02: Angular measurement 0x04: Digital inputs 0x08: Environmental compensation
CollectDataUpdateTimeScalar: DevULong Logarithmic time distance of two subsequent position measurements. The actual sample time can be varied only in powers of two: sample time = 10.24us * (2 ^ lbSmpTime). Allowed range is 0 ... 24 (10.24us ... 172s) . The sample time directly affects the data rate!
CollectDataMaxSizeScalar: DevULong The maximum number of position data elements the CollectDataStart / CollectDataTime are allowed to collect. Once this value is exceeded data collection will be stopped. This is just a precaution tp prevent the program from eating up all the memory if someone forgets to call CollectDataStop or runs CollectDataTime with a too high setting.
FileDirScalar: DevString
FilePrefixScalar: DevString
FileStartNumScalar: DevLong
FilePostfixScalar: DevString
FileSavingScalar: DevBoolean
TemperatureScalar: DevDouble
AirPressureScalar: DevDouble
HumidityScalar: DevDouble
RefractionScalar: DevDouble
PositionsSpectrum: DevDouble

Commands:

Name Description
ResetAxisInput: DevULong
axis number (0-2)
Output: DevVoid
SignalQualityInput: DevULong
Axis number (0-2)
Output: DevULong
signal quality (the higher the better)
StateInput: DevVoid
Output: State
Device state
This command gets the device state (stored in its device_state data member) and returns it to the caller.
StatusInput: DevVoid
Output: ConstDevString
Device status
This command gets the device status (stored in its device_status data member) and returns it to the caller.
AxisStatusInput: DevULong
axis number
Output: DevLong
0: axis is successfully aligned, signal quality is sufficient 1: axis is successfully aligned, signal quality is insufficient 2: axis is not successfully aligned, signal quality is sufficient 3: axis is not successfully aligned, signal quality is insufficient
AdjustInput: DevVoid
Output: DevVoid
start the adjustment procedure
RecalibrateInput: DevVoid
Output: DevVoid
Start the recalibration procedure.
SetPollTimeInput: DevVarULongArray
arg[0] : axis number arg[1] : poll time (= 80ns * (2 ^ arg[1]), range 0->31
Output: DevVoid
GetPollTimeInput: DevULong
axis number
Output: DevULong
poll time (= 80ns * (2 ^ arg[1]), range 0->31
get poll time for axis
CollectDataTimeInput: DevULong
time to run position data collection in ms
Output: DevVoid
time to run position data collection in ms
GetMeanDevInput: DevULong
Output: DevVarDoubleArray
0: average value 1: standard deviation
Get results per axis after a CollectData call
RunCallBackInput: DevVarUShortArray
0:true: start callback function 0:false: stop callback function 1:true routine ran ok 1:false: routine hit datasize limit
Output: DevVoid
for internal use only, never call from client
CollectDataStartInput: DevVoid
start position data collection
Output: DevVoid
CollectDataStopInput: DevVoid
stop position data collection
Output: DevVoid
GetCollectDataInput: DevULong
Axis number
Output: DevVarDoubleArray

Pipes:

Properties:

Name Description
DeviceIDDevLong ID of device, if no IpAdress is set or ConnectType is empty or set to `usb`, the ID is used to identify the device.
ConnectTypeDevString Connection method, may be `net` for tcp/ip or `usb` for usb. If empty / not set, devices will be searched on both types.
HostNameDevString IP Address or hostname If set, device will be identified by this name/address, otherwise identification will be done by DeviceID.
TriggerSourceDevString Device/Attribute which emits trigger signal to start data collection
UseTriggerDevBoolean whether to use trigger (true) or not (false)

23 Feb 2017, Piotr Goryl
Updated:
The device class has been updated.
You can see previous version here .



19 Jan 2017, Piotr Goryl
Updated:
The device server has been updated.
You are looking at this version now.



2 Jan 2017, Piotr Goryl
Created:
The device server has been added to catalogue.
Added by:pgoryl2 on:19 Jan 2017, 10:42 a.m.