Class Description


Server for the controll of the Pressuer Automated Calibration Equipment 5000 from GE

Families: MeasureInstruments

Key words:

Platform: Unix Like

Language: Python

License: GPL

Contact:

Hardware


Manufacturer: GE

Product:

PACE-5000

Bus: Socket

Class interface


Attributes:

Name Description
MembranePressureScalar: DevDouble Current output pressure
InletPressureScalar: DevDouble Current input pressure
SetPointScalar: DevDouble Setpoint output line pressure
SlewRateScalar: DevDouble Slew rate for the output line pressure control
SlewRateModeScalar: DevBoolean Instrumental mode for the slew rate operation – controls usage of the SlewRate parameter. Maximum slew rate – SlewRate parameter is disregarded and compression or decompression takes place with maximum pace. Linear – device respects the SlewRate setting and performs the compression or decompression accordingly
OvershootScalar: DevBoolean Parameter controlling the overshooting capability on pressure increase: 0 – not allowed 1 – allowed
PressureControlScalar: DevBoolean Parameter controlling the main operation of the pump. If pressure control is ON – the device does it best to increase or decrease the pressure according to the setpoint. f pressure control is OFF – the device is going to the measurement mode, no control, just measurement (useful for leakage testing)
SourceEffectivenessScalar: DevDouble Parameter reflecting effectiveness of the effort in order to achieve the setpoint
DecompressRateScalar: DevDouble Decompression rate set by the user. Used in conjunction with the Decompress Tango Device Command

Commands:

Name Description
StateInput: DevVoid
Output: State
Device state
This command gets the device state (stored in its device_state data member) and returns it to the caller.
StatusInput: DevVoid
Output: ConstDevString
Device status
This command gets the device status (stored in its device_status data member) and returns it to the caller.
StopInput: DevVoid
Output: DevVoid
Stops operation of the pump. The most important action – stop the PressureControl – power control – go to the measurement mode
DecompressInput: DevVoid
Output: DevVoid
Starts the decompression to the lowest pressure with DecompressRate First – stops the operation of the pump. Reads the value of the DecompressRate, sets the value of the SlewRate to this value Sets the SlewRateMode to 0=LINEAR Sets the SetPoint = 0 Sets the PressureControl = 1
CompressInput: DevVoid
Output: DevVoid
Starts the pump to do compression or decompression PressureControl = 1
GetIDNInput: DevVoid
Output: DevString
Returns the identification string from the device as a string
GetMACInput: DevVoid
Output: DevString
Returns the MAC address from the device as a string
GetErrorStateInput: DevVoid
Output: DevString
Returns the error message from the instrument
ClearErrorStateInput: DevVoid
Output: DevVoid
Clears the error state for the instrument. This is done by getting the GetErrorState – 5x times to clear the inner error message buffer
MakeHWTestInput: DevVoid
Output: DevLong
1-> OK, 0-> Fail
Performs a hardware test for the instrument. If the result is positive – device goes to the ON state, else, the device goes to the ALARM state and disallow any SCPI command to go through.
SendCmdInput: DevString
Command to send
Output: DevString
Return value in case it exists, empty string in other case.
Send command to pressure controller and return the answer in case it exits.

Pipes:

Properties:

Name Description
IPAdressDevString IP Adress
PortNbDevLong Port Number

20 Feb 2021, DS Admin
Updated:
The device class has been updated.
You are looking at this version now.



2 Feb 2019, DS Admin
Updated:
The device class has been updated.
You can see previous version here .



23 Feb 2017, Piotr Goryl
Created:
The device class has been added to catalogue.
Added by:pgoryl2 on:20 Feb 2021, 3:39 a.m.